Image quality and cancer visibility of T2-weighted magnetic resonance imaging of the prostate at 7 Tesla

PMID: PMID
DOI: DOI
Journal: European Radiology
Year of publication: 2014
Page: 24(8):1950-8

E.K. Vos, M.W. Lagemaat, J.O. Barentsz, J.J. Fütterer, P. Zámecnik, H. Roozen, S. Orzada, A.K. Bitz, M.C. Maas & T.W.J. Scheenen

OBJECTIVES: To assess the image quality of T2-weighted (T2w) magnetic resonance imaging of the prostate and the visibility of prostate cancer at 7 Tesla (T).

MATERIALS AND METHODS: Seventeen prostate cancer patients underwent T2w imaging at 7T with only an external transmit/receive array coil. Three radiologists independently scored images for image quality, visibility of anatomical structures, and presence of artefacts. Krippendorff's alpha and weighted kappa statistics were used to assess inter-observer agreement. Visibility of prostate cancer lesions was assessed by directly linking the T2w images to the confirmed location of prostate cancer on histopathology.

RESULTS: T2w imaging at 7T was achievable with 'satisfactory' (3/5) to 'good' (4/5) quality. Visibility of anatomical structures was predominantly scored as 'satisfactory' (3/5) and 'good' (4/5). If artefacts were present, they were mostly motion artefacts and, to a lesser extent, aliasing artefacts and noise. Krippendorff's analysis revealed an α = 0.44 between three readers for the overall image quality scores. Clinically significant cancer lesions in both peripheral zone and transition zone were visible at 7T.

CONCLUSION: T2w imaging with satisfactory to good quality can be routinely acquired, and cancer lesions were visible in patients with prostate cancer at 7T using only an external transmit/receive body array coil.