Simultaneous R2*, R2, and R2' quantification by combining S0 estimation of the free induction decay with a single spin echo: A single acquisition method for R2 insensitive quantification of holmium-166-loaded microspheres

PMID: PMID
DOI: DOI
Journal: Magnetic Resonance in Medicine
Year of publication: 2015
Page: 73(1):273-83

G.H. van de Maat, H. de Leeuw, P.R. Seevinck, M.A.A.J. van den Bosch, J.F.W. Nijsen & C.J.G. Bakker

PURPOSE: To present a new method, S0 estimation of the free induction decay combined with a single spin echo measurement (SOFIDSE), that enables simultaneous measurements of R2*, R2, and R2' in order to quantify the local concentration of holmium microspheres (Ho-MS) for radioembolization.

THEORY AND METHODS: SOFIDSE estimates R2* and the signal magnitude at time point 0, S0, from a multigradient echo readout of the free induction decay and subsequently estimates R2 using S0 and a single spin echo, from which R2' is deducted. The method was evaluated by comparing SOFIDSE R2 values with values obtained from shifted spin echo (SSE) measurements on a phantom setup containing Ho-MS and from dual spin echo measurements on a healthy volunteer.

RESULTS: On average, SOFIDSE showed a small overestimation of R2 values compared with SSE independent of the microsphere concentration. R2' values determined by subtraction of either SOFIDSE R2 or SSE R2 from R2* showed excellent agreement (correlation coefficient = 1; P = 9 · 10(-11)). The Ho-MS-induced R2' values obtained by SOFIDSE were insensitive to the R2 value of the tissue in which they resided.

CONCLUSION: SOFIDSE enables quantification of Ho-MS, in media with spatially or temporally varying R2 values, in a single acquisition.